ryan.crouch@LIGO.ORG - posted 12:10, Monday 05 August 2024 - last comment - 10:26, Tuesday 06 August 2024(79448)
ETMX OPLEV charge measurement

I took the ETMX OPLEV charge measurement, it's been about a week since Rahul updated the L3 HV mon alog79350.

The charge appears to be flatlining now, and even just starting to decrease towards zero on some quads/DOF (UR_Y, LR, LL_P).

Images attached to this report
Comments related to this report
ryan.crouch@LIGO.ORG - 10:26, Tuesday 06 August 2024 (79472)CDS

The last 2 times I've run the OPLEV charge measurement for ETMX, about an hour later we've had a power glitch that takes down the VFD at the end station, 08/05 alog79455, 07/23 alog79288. Weird coincidence?