Pressure Gauge Failure At The Filter Cavity Tube Location A2

Yesterday PT154 was found with a blinking "amber" LED, per the user manual:

Blinking Amber 1. Both Micro-Ion gauge filaments inoperable – Replace the gauge
2. Possible electronics failure – Return the module to MKS

Gauge will stay there for now, if posible and approved by others a replacement will be installed.  A plot of the failure is attached, signal for the pressure gauge is the RED trace.

Unfortunately this is not the only one that I found that is in trouble, another gauge located at position C1 has a continuous "amber" LED, per the user manual:

Steady Amber Micro-Ion gauge ON – One filament is inoperable

I walked and visited all other gauges, MKS brand, and they appear to be OK.

 

FRS ticked opened for faulty gauge, 31787.