Displaying report 1-1 of 1.
Reports until 17:38, Thursday 22 August 2024
H1 SQZ
camilla.compton@LIGO.ORG - posted 17:38, Thursday 22 August 2024 (79651)
SQZ Alignment using SQZ Single Bounce

Naoki, Vicky, Camilla, Sheila. Copied instructions from 78115.

Today we couldn't find the SQZ beam while the IFO was locked using  SQZ-OMC_TRANS_RF3_DEMOD_RFMON as we'd been ale to do in 78088. When the IFO unlocked we used the SQZ SEED beam in a SQZ single bounce (much higher power so easier to see) to  improve the SQZ alignment. Expect when we're locked will still need to touch alignment, may get away with a few rounds of SQZ_MANAGER's SCAN_ALIGNMENT_FDS state (wiki notes). Notes below.

Images attached to this report