corey.gray@LIGO.ORG - posted 16:29, Saturday 24 August 2024 - last comment - 08:17, Monday 26 August 2024(79683)
Sat DAY Ops Summary

TITLE: 08/24 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Lock Acquisition
INCOMING OPERATOR: Oli
SHIFT SUMMARY:

Have had the BS camera die on us a couple times in the last 24hrs (this requires contacting Dave to restart the camera's computer--atleast until this camera can be moved to another computer), and so the beginning of the shift was restoring this and also figuring out why H1 had a lockloss, BUT did not drop the Input Power down from 61W to 2W.

After the camera issue was fixed, ran an alignment with no issues, and then took H1 all the back to NLN also with no issues.  

Then completed taking Calibration measurements after 1.5hrs and 3hrs.  Oli also ran LSC Feedforward measurements......

Then there was ANOTHER BS Camera Computer Crash!  Dave brought us back fast.  

Now back to OBSERVING!
LOG:

Comments related to this report
ryan.short@LIGO.ORG - 08:17, Monday 26 August 2024 (79707)ISC, OpsInfo

Looking into why the input power didn't come back down after a lockloss at ADS_TO_CAMERAS, seems that the proper decorators that check for a lockloss are missing from the run method (but are there in main). This means that while ISC_LOCK was waiting for the camera servos to turn on, it didn't notice that the IFO lost lock, and therefore didn't run through the LOCKLOSS or DOWN states which would reset the input power.

I've added the decorators to the run method of ADS_TO_CAMEARS so this shouldn't happen again.

 