PCALY SDF diffs

There were a few PCAL SDF diffs when we made it back to low noise today at CALEY. It looked like these were loaded into epics from teh safe.snap and these numbers disagreed with what was saved in the observe.snap file. I confirmed with Franscisco that this was the case, and he had me verify that they agreed with alog77386. I then also saved these new values in the safe.snap file as well. Interestingly, these channels are not monitored in the safe.snap.