Displaying report 1-1 of 1.
Reports until 15:39, Wednesday 28 August 2024
H1 SUS
ryan.crouch@LIGO.ORG - posted 15:39, Wednesday 28 August 2024 (79773)
Weekly Inlock charge measurement

Closes FAMIS 28368
 

The processing and plotting code were both giving me some trouble, from not creating the coefficient files in /opt/rtcds/userapps/release/sus/common/scripts/quad/InLockChargeMeasurements/rec_LHO_charge_coeff/, and a bad ITMX measurement from July making shapes not match.

ITMXs measurement is also possibly not good, as it was complaining about its data structure and I was not able to process it.

raise ValueError("no Fr{{Adc,Proc,Sim}}Data structures with the "
ValueError: no Fr{Adc,Proc,Sim}Data structures with the name H1:SUS-ITMX_L3_ESDAMON_DC_OUT_DQ


The charges seem stagnant

Images attached to this report
Displaying report 1-1 of 1.