Displaying report 1-1 of 1.
Reports until 06:57, Saturday 31 August 2024
H1 SQZ
oli.patane@LIGO.ORG - posted 06:57, Saturday 31 August 2024 - last comment - 10:36, Saturday 31 August 2024(79833)
SQZ OWL Intervention

Had to help H1 Manager out because we were in NOMINAL_LOW_NOISE but the OPO was having trouble getting locked with the ISS. OPO trans was having trouble getting to 70 and so definitely couldn't reach its setpoint of 80uW. I changed the setpoint to 69 since it was maxing out around 69.5, and I changed the OPO temp and accepted it in sdf.

Images attached to this report
Comments related to this report
naoki.aritomi@LIGO.ORG - 10:36, Saturday 31 August 2024 (79839)

We had some SHG rejected power so I rotated HWP to minimize the SHG rejected power and get more SHG launch power. Then I reverted the OPO trans setpoint to 80 and reverted the OPO temperature as shown in the attached SDF. 

Images attached to this comment
Displaying report 1-1 of 1.