The following gains were set to Zero:
caput H1:CAL-PCALY_PCALOSC1_OSC_SINGAIN 0
caput H1:CAL-PCALY_PCALOSC2_OSC_SINGAIN 0
caput H1:CAL-PCALY_PCALOSC3_OSC_SINGAIN 0
caput H1:CAL-PCALY_PCALOSC4_OSC_SINGAIN 0
caput H1:CAL-PCALY_PCALOSC9_OSC_SINGAIN 0
caput H1:CAL-PCALX_PCALOSC1_OSC_SINGAIN 0
caput H1:CAL-PCALX_PCALOSC1_OSC_COSGAIN 0
caput H1:CAL-PCALX_PCALOSC4_OSC_SINGAIN 0
caput H1:CAL-PCALX_PCALOSC5_OSC_SINGAIN 0
caput H1:CAL-PCALX_PCALOSC6_OSC_SINGAIN 0
caput H1:CAL-PCALX_PCALOSC7_OSC_SINGAIN 0
caput H1:CAL-PCALX_PCALOSC8_OSC_SINGAIN 0
Then I took ISC_LOCK to NLN_CAL_MEAS
22:13 UTC ran the calibration following command:
pydarm measure --run-headless bb
notification: end of test
diag> save /ligo/groups/cal/H1/measurements/PCALY2DARM_BB/PCALY2DARM_BB_20240831T221301Z.xml
/ligo/groups/cal/H1/measurements/PCALY2DARM_BB/PCALY2DARM_BB_20240831T221301Z.xml saved
diag> quit
EXIT KERNEL
2024-08-31 15:18:13,008 bb measurement complete.
2024-08-31 15:18:13,008 bb output: /ligo/groups/cal/H1/measurements/PCALY2DARM_BB/PCALY2DARM_BB_20240831T221301Z.xml
2024-08-31 15:18:13,008 all measurements complete.
anthony.sanchez@cdsws29:
anthony.sanchez@cdsws29: gpstime;python /ligo/groups/cal/src/simulines/simulines/simuLines.py -i /ligo/groups/cal/src/simulines/simulines/newDARM_20231221/settings_h1_newDARM_scaled_by_drivealign_20231221_factor_p1.ini;gpstime
PDT: 2024-08-31 15:18:49.210990 PDT
UTC: 2024-08-31 22:18:49.210990 UTC
GPS: 1409177947.210990
2024-08-31 22:41:50,395 | INFO | Commencing data processing.
Traceback (most recent call last):
File "/ligo/groups/cal/src/simulines/simulines/simuLines.py", line 712, in
run(args.inputFile, args.outPath, args.record)
File "/ligo/groups/cal/src/simulines/simulines/simuLines.py", line 205, in run
digestedObj[scan] = digestData(results[scan], data)
File "/ligo/groups/cal/src/simulines/simulines/simuLines.py", line 621, in digestData
coh = np.float64( cohArray[index] )
File "/var/opt/conda/base/envs/cds/lib/python3.10/site-packages/gwpy/types/series.py", line 609, in __getitem__
new = super().__getitem__(item)
File "/var/opt/conda/base/envs/cds/lib/python3.10/site-packages/gwpy/types/array.py", line 199, in __getitem__
new = super().__getitem__(item)
File "/var/opt/conda/base/envs/cds/lib/python3.10/site-packages/astropy/units/quantity.py", line 1302, in __getitem__
out = super().__getitem__(key)
IndexError: index 3074 is out of bounds for axis 0 with size 0
ICE default IO error handler doing an exit(), pid = 2858202, errno = 32
PDT: 2024-08-31 15:41:53.067044 PDT
UTC: 2024-08-31 22:41:53.067044 UTC
GPS: 1409179331.067044
These changes were reverted and restored back to their previous values.
H1:CAL-PCALY_PCALOSC1_OSC_SINGAIN
H1:CAL-PCALY_PCALOSC2_OSC_SINGAIN
H1:CAL-PCALY_PCALOSC3_OSC_SINGAIN
H1:CAL-PCALY_PCALOSC4_OSC_SINGAIN
H1:CAL-PCALY_PCALOSC9_OSC_SINGAIN
H1:CAL-PCALX_PCALOSC1_OSC_SINGAIN
H1:CAL-PCALX_PCALOSC1_OSC_COSGAIN
H1:CAL-PCALX_PCALOSC4_OSC_SINGAIN
H1:CAL-PCALX_PCALOSC5_OSC_SINGAIN
H1:CAL-PCALX_PCALOSC6_OSC_SINGAIN
H1:CAL-PCALX_PCALOSC7_OSC_SINGAIN
H1:CAL-PCALX_PCALOSC8_OSC_SINGAIN
I then took ICS_LOCK back back to NOMINAL_LOW_NOISE.