This icalibration was requested by Sheila after today's Commissioning work.
Measurement NOTES:
- Following instructions (link)
- H1 at NLN for for over 13.25hrs
- Calibration_Monitor medm screenshot is attached
- 1827utc (1409596056 gps) Out of Observing, ISC LOCK to NLN_CAL_MEAS & Observatory Mode to CALIBRATION
- 1827-1832 Broadband measurement run
- Measurement at: /ligo/groups/cal/H1/measurements/PCALY2DARM_BB/
- 1832-1856 Simullines measurement
- gps START time: 1409596387.219105
- gps END time: 1409597812.855494
- Files writtten to subfolders here: /ligo/groups/cal/H1/measurements