This is the 3rd calibration of the day. The one taken post-commissioning was not usable/had error. So received request to run another Calibration.
Measurement NOTES:
- Following instructions (link)
- H1 at NLN for for over 13.25hrs
- Calibration_Monitor medm screenshot is attached
- 2003utc (1409596056 gps) Out of Observing, ISC LOCK to NLN_CAL_MEAS & Observatory Mode to CALIBRATION
- 2003-2008 Broadband measurement run
- Measurement at: /ligo/groups/cal/H1/measurements/PCALY2DARM_BB/
- 2008-2031 Simullines measurement
- gps START time: 1409602151.842028
- gps END time:1409603535.426098
- Files writtten to subfolders here: /ligo/groups/cal/H1/measurements