Thomas, Mitchell & Jason got the majority of the fixturing out of the chamber, bagged & wrapped, and out of the VEA. They then got the retroreflector hanging on the Quad. We then unlocked the HEPI and leveled & set the elevation of the Optical Table per D0901152. I set the system ~.5mm low as IAS found the Optic 0.4mm high on the Test Stand. The level is within +-0.1mm and the elevation is within 0.3mm; good enough since we still likely have much horizontal adjustment to do. With the retroreflector in place, we then shot an X position. IAS is still evaluating that result.