Displaying report 1-1 of 1.
Reports until 16:15, Tuesday 10 September 2024
H1 DAQ
daniel.sigg@LIGO.ORG - posted 16:15, Tuesday 10 September 2024 - last comment - 15:53, Thursday 12 September 2024(80023)
New DAC gain and delay measurements

Marc Daniel

We measured the gain and phase difference between the new DAC and the existing 20-bit DAC in SUS ETMX. For this we injected 1kHz sine waves and measure gain and phase shifts between the two. We started with a digital gain value of 275.65 for the new DAC and adjusted it to 275.31 after the measurement to keep the gains identical. The new DAC implements a digital AI filter that has a gain of 1.00074 and a phase of -5.48° at 1kHz, which corresponds to a delay of 15.2µs.

This puts the relative gain (new/old) to 1.00074±0.00125 and the delay to 13.71±0.66µs. The variations can be due to the gain variations in the LIGO DAC, the 20-bit DAC, the ADC or the AA chassis.

DAC Channel Name Gain Adjusted Diff (%) Phase (°) Delay (us)
0 H1:SUS-ETMX_L3_ESD_DC 1.00239 1.00114 0.11% -5.29955 -14.72
1 H1:SUS-ETMX_L3_ESD_UR 1.00026 0.99901 -0.10% -5.10734 -14.19
2 H1:SUS-ETMX_L3_ESD_LR 1.00000 0.99875 -0.12% -4.93122 -13.70
3 H1:SUS-ETMX_L3_ESD_UL 1.00103 0.99978 -0.02% -5.11118 -14.20
4 H1:SUS-ETMX_L3_ESD_LL 1.00088 0.99963 -0.04% -5.21524 -14.49
8 H1:SUS-ETMX_L1_COIL_UL 1.00400 1.00275 0.27% -4.72888 -13.14
9 H1:SUS-ETMX_L1_COIL_LL 1.00295 1.00170 0.17% -4.88883 -13.58
10 H1:SUS-ETMX_L1_COIL_UR 1.00125 1.00000 0.00% -5.08727 -14.13
11 H1:SUS-ETMX_L1_COIL_LR 1.00224 1.00099 0.10% -4.92882 -13.69
12 H1:SUS-ETMX_L2_COIL_UL 1.00325 1.00200 0.20% -4.78859 -13.30
13 H1:SUS-ETMX_L2_COIL_LL 1.00245 1.00120 0.12% -4.55283 -12.65
14 H1:SUS-ETMX_L2_COIL_UR 1.00175 1.00050 0.05% -4.52503 -12.57
15 H1:SUS-ETMX_L2_COIL_LR 1.00344 1.00219 0.22% -5.00466 -13.90
  Average 1.00199 1.00074 0.07% -4.93611 -13.71
  Standard Deviation 0.00125 0.00125 0.13% 0.23812 0.66
Comments related to this report
daniel.sigg@LIGO.ORG - 15:53, Thursday 12 September 2024 (80064)

FPGA filter is

zpk([585.714+i*32794.8;585.714-i*32794.8;1489.45+i*65519.1;1489.45-i*65519.1;3276.8+i*131031; \
    3276.8-i*131031;8738.13+i*261998;8738.13-i*261998], \
    [11555.6+i*17294.8;11555.6-i*17294.8;2061.54+i*26720.6;2061.54-i*26720.6;75000+i*93675; \
    75000-i*93675;150000+i*187350;150000-i*187350;40000],1,"n")

Non-image files attached to this comment
Displaying report 1-1 of 1.