Sheila and I are still trying to determine if the a2l script can be helpful, or made helpful, and understand our coupling from these changes to DHARD and CHARD. We ran it today with both C&D HARD showing higher coherences with our current lock (attachment from Ryan C's alog80244). After running the script the largest change was to ITMY Y gain (-0.28 diff). This helped out DHARD from 15-35Hz, but made CHARD worse in 10-21Hz. We were then going to try and tune common by hand to see if we could make any difference, but we lost lock.
We decided to update the gains with the script results since this did help out our DHARD. Perhaps an acceptable trade off for now. Values for the Y gains have been updated in lscparams and SDF, but P was close enough that I left them alone.