Displaying report 1-1 of 1.
Reports until 13:37, Tuesday 24 September 2024
H1 CAL
louis.dartez@LIGO.ORG - posted 13:37, Tuesday 24 September 2024 - last comment - 14:37, Tuesday 24 September 2024(80270)
calibration updated
During today's maint. period I updated the calibration on CAL-CS and restarted the GDS pipeline. The kappas will be reset to 1.

The biggest reason for this update is to account for uncompensated delays left over from the DAC swap (& others that have been in place but that we just never accounted for).

new pydarm H1 ini file
latest pydarm report
Images attached to this report
Comments related to this report
david.barker@LIGO.ORG - 14:29, Tuesday 24 September 2024 (80273)

Related to this the CDS SDF, which is monitoring the CAL_HASH ioc is reporting three differences

Images attached to this comment
louis.dartez@LIGO.ORG - 14:37, Tuesday 24 September 2024 (80274)
Thanks Dave. I've accepted those now too.