H1 SUS
arnaud.pele@LIGO.ORG - posted 20:14, Friday 04 October 2013 - last comment - 14:41, Tuesday 08 October 2013(8007)
OLV RM1 RM2

Open light values have been measured on RM1 and RM2

OPTIC OSEM Open Ligh Value (cts) Gain Offset Serial Number
RM1 M1UL -32768 -0.916 16384 1105240 a
M1LL -32768 -0.916 16384 1105240 b
M1UR -22268 -1.347 11134 1105240 c
M1LR -29171 -1.028 14586 1105240 d
RM2 M1UL -26656 -1.125 13328 1105238 a
M1LL -32768 -0.916 16384 1105238 b
M1UR -32708 -0.917 16354 1105238 c
M1LR -32768 -0.916 16384 1105238 d

damping gains have been set to

-30 in long

-0.02 in Pitch

-0.02 in Yaw

the safe snapshot has been commited to the svn

Comments related to this report
arnaud.pele@LIGO.ORG - 14:41, Tuesday 08 October 2013 (8043)

I used the script "[OSEM2EUL, EUL2OSEM, SENSALIGN, DRIVEALIGN] = make_sushtts_projections" in /HTTS/Common/MatlabTools/ to generate the matrices for RM1 and RM2

They have been automatically filled with fill_matrix_values.m

The values are

OSEM2EUL:

    0.2500    0.2500    0.2500    0.2500
   10.3681  -10.3681   10.3681  -10.3681
  -10.3681  -10.3681   10.3681   10.3681

EUL2OSEM matrix :

    0.2500   10.3681  -10.3681
    0.2500  -10.3681  -10.3681
    0.2500   10.3681   10.3681
    0.2500  -10.3681   10.3681

Safe snapshot has been saved for both RM1 and RM2