Reports until 13:51, Monday 07 October 2024
H1 CDS
david.barker@LIGO.ORG - posted 13:51, Monday 07 October 2024 - last comment - 13:58, Monday 07 October 2024(80511)
VACSTAT glitch alarm for BSC3 (false alarm)

We had another transient glitch of the BSC3 PT132-MOD2 gauge which triggered the VACSTAT alarms. Gerardo confirmed that no other gauge had an issue.

The PT132 glitch is a square wave, only 2 seconds in width.

Pressure went from 4.8e-08 to 6.7e-08 Torr at 13:25:42 and returned to its base value 2 seconds later.

VACSTAT MEDM and comparison between BSC2 and BSC3 attached.

Images attached to this report
Comments related to this report
david.barker@LIGO.ORG - 13:58, Monday 07 October 2024 (80512)

I restarted vacstat_ioc.service on cdsioc0 at 13:53 to clear this BSC3 latched event.