Displaying report 1-1 of 1.
Reports until 08:36, Thursday 10 October 2024
H1 SQZ
camilla.compton@LIGO.ORG - posted 08:36, Thursday 10 October 2024 (80586)
Lockloss before Squeezing Data Set Taken

Sheila, Camilla

15:02:40-15:13:00UTC - Took No Squeezing time. IFO had been in NLN for 4h30m.

Checked NLG (0.1178/0.00872) = 13.5. Adjusted OPO temperature. 

After our no SQZ time, once in FDS, we let the IFO ASC converge and then turned it off, cleared histories in SQZ-ASC filter banks and ran scan sqz alignment. We never checked these were the same in 80506. Lost lock before scan sqz alignment could finish running. 