Calibration measurement run after being locked for 3 hours
calibration monitor screenshot (accidently taken after going into NLN_CAL_MEAS)
Broadband 10/10 21:50 - 21:56UTC
File:
PCALY2DARM_BB/PCALY2DARM_BB_20241010T215057Z.xml
Simulines* 10/10 21:57 - 22:21UTC
*Simulines was run using the ini file settings_h1_20241005_lowerPcal.ini instead of the usual
settings_h1_newDARM_scaled_by_drivealign_20231221_factor_p1.ini, both of which are found in /ligo/groups/cal/H1/simulines_settings/newDARM_20231221/
Files:
DARMOLG_SS/DARMOLG_SS_20241010T215755Z.hdf5
PCALY2DARM_SS/PCALY2DARM_SS_20241010T215755Z.hdf5
SUSETMX_L1_SS/SUSETMX_L1_SS_20241010T215755Z.hdf5
SUSETMX_L2_SS/SUSETMX_L2_SS_20241010T215755Z.hdf5
SUSETMX_L3_SS/SUSETMX_L3_SS_20241010T215755Z.hdf5