Displaying report 1-1 of 1.
Reports until 12:42, Wednesday 16 October 2024
H1 TCS
jackie.figueroa@LIGO.ORG - posted 12:42, Wednesday 16 October 2024 (80714)
Absorption measurement for lockloss time 1412435458.
From Camilla and Jackie

Last done in April 2024

plots for reference, 1412435458.png, HWS_aborbs.png

Using  /ligo/gitcommon/labutils/hws_absorption_fit/april2024/fit_absprtion_v2.py with lockloss time 1412435458, edited P_y  and P_x values to adjust the fit.

ITMX: 155mW absorbed = 398ppb (155mW / 389kW) 

ITMY: 140mW absorbed = 360ppb (140mW / 389kW)

  April 2022 (6246862782) Nov 2022 (66036 April 2023 (71400) April 2024 October 2024
ITMX 430ppb 490ppb 475ppb 430ppb

398ppb

ITMY 370ppb 385ppb 375ppb 330ppb

360ppb

Images attached to this report
Displaying report 1-1 of 1.