Displaying report 1-1 of 1.
Reports until 09:14, Friday 18 October 2024
H1 CDS
david.barker@LIGO.ORG - posted 09:14, Friday 18 October 2024 - last comment - 10:05, Saturday 19 October 2024(80742)
CNS-II reference GPS at EY, another -800nS glitch

From 07:32 - 07:37 Fri 18oct2024 PDT we had another glitch on the EY CNS-II GPS 1PPS as read by the timing system's EY comparator. This had happened twice recently, 05:50-06:40 Mon 30sep2024 (50mins) and a pair the next day 07:14 (3mins) 07:42 (8mins) Tue 01oct2024.

attached are details of today's glitch, and a month trend showing all three.

On a related note, the EX CNS-II broke on the 8th Oct with a bad power supply (a wall wart). It might be prudent to schedule a replacement of EY's power supply next Tuesday.

Images attached to this report
Comments related to this report
david.barker@LIGO.ORG - 10:16, Friday 18 October 2024 (80743)

It started again at 10:02. This time it is switching between -2200ns and -800ns

Images attached to this comment
david.barker@LIGO.ORG - 10:05, Saturday 19 October 2024 (80757)

Follow up: EY CNS-II started glitching again around 10am Friday, it continued to do this for about an hour, then went good again. No further glitches since then over the past 23 hours.

Displaying report 1-1 of 1.