Reports until 16:59, Monday 21 October 2024
H1 ISC
camilla.compton@LIGO.ORG - posted 16:59, Monday 21 October 2024 - last comment - 11:06, Thursday 09 January 2025(80804)
Automatic DARM Offset Steps

Elenna, Camilla

Ran the automatic DARM offset sweep via Elenna's instructions (took <15 minutes):

cd /ligo/gitcommon/labutils/darm_offset_step/
conda activate labutils
python auto_darm_offset_step.py

DARM offset moves recorded to /ligo/gitcommon/labutils/darm_offset_step/data/darm_offset_steps_2024_Oct_21_23_37_44_UTC.txt

Comments related to this report
camilla.compton@LIGO.ORG - 17:02, Monday 21 October 2024 (80805)

Reverted the attached tramp sdf diffs afterwards, see attached. Maybe the script needs to be adjusted to automatically do this.

Images attached to this comment
elenna.capote@LIGO.ORG - 17:28, Monday 21 October 2024 (80806)

I just ran Craig's script to analyze these results. The script fits a contrast defect of 0.742 mW using the 255.0 Hz data and 0.771 mW using the 410.3 Hz data. This value is lower than the previously 1 mW on July 11 (alog 79045), which matches up nicely with our reduced frequency noise since the OFI repair (alog 80596).

I attached the plots that the code generates.

This result then estimates that the homodyne angle is about 7 degrees.

Non-image files attached to this comment
jennifer.wright@LIGO.ORG - 11:06, Thursday 09 January 2025 (82204)

Last year I added some code to plot_darm_optical_gain_vs_dcpd_sum.py to calculate the losses through HAM 6.

Sheila and I have started looking at those again post-OFI replacement.

Just attaching the plot of power at the anti-symmetric port (ie. into HAM 6) vs. power after the OMC as measured by the DCPDs.

The plot is found in /ligo/gitcommon/labutils/darm_offset_step/figures/ and is also on the last page of the pdf Elenna linked above.

From this plot we can see that the relationship between the power into HAM 6 (P_AS) is related to the power at the output DCPDs as follows.

P_AS = 1.220*P_DCPD + 656.818 mW

Where the second term is light that will be rejected by the OMC + that which gets through the OMC but is insensitive to DARM length changes.

The throughput between the anti-symmetric port and the DCPDs is 1/1.22 = 0.820. So that means 18% of the TM00 light that we want at the DCPDs is lost through HAM 6.

 

Non-image files attached to this comment