Displaying report 1-1 of 1.
Reports until 11:15, Friday 01 November 2024
H1 ISC
sheila.dwyer@LIGO.ORG - posted 11:15, Friday 01 November 2024 - last comment - 15:47, Friday 01 November 2024(80993)
FSS OLG measurement

We measured the TTFSS OLG, I didn't save the data but will attach several photos here. There is a feature that looks like it could be unstable at 1.68MHz, there are about 3 untiy gain crossings around 500 kHz.  The FSS has a gain between the test points that mean that unity gain is at -10dB in this measurement. 

67253 Craig mentions that the FSS OLG was 350kHz, we'd need to reduce the gain by 3dB to get back to a 350 kHz ugf.

Comments related to this report
sheila.dwyer@LIGO.ORG - 11:17, Friday 01 November 2024 (80994)
Images attached to this comment
marc.pirello@LIGO.ORG - 14:43, Friday 01 November 2024 (80996)

Scans from the FSS after reducing gain.

Non-image files attached to this comment
sheila.dwyer@LIGO.ORG - 15:47, Friday 01 November 2024 (80998)

In Marc's files above the A one should be magnitude for 100kHz to a few MHz frequency span, B should be phase for that measurement.  C should be magntiude zoomed in on the 1.7MHz region, D should be phase for that measurement. 

Images attached to this comment
Displaying report 1-1 of 1.