Vicky, Camilla
I've updated instructions in the SQZ wiki: https://cdswiki.ligo-wa.caltech.edu/wiki/Troubleshooting%20SQZ
If the squeezing runs away, you should now be able to take SQZ_MANAGER to '''RESET_SQZ_ASC_FDS''' (turns off and clear the ASC and turns off the angle servo, will also reset the SQZ angle to 190 deg if it's outside 100-250 range) and the request FREQ_DEP_SQZ (will got through SQZ_ASC_FDS) to turn back on SQZ ASC and angle control.
This is to avoid issues as in 81027. We still might need to think about what should happen to the FC ASC.