Closes FAMIS#26316, last checked 80936
Laser Status:
NPRO output power is 1.9W (nominal ~2W)
AMP1 output power is 68.62W (nominal ~70W)
AMP2 output power is 138.7W (nominal 135-140W)
NPRO watchdog is GREEN
AMP1 watchdog is GREEN
AMP2 watchdog is GREEN
PDWD watchdog is GREEN
PMC:
It has been locked 0 days, 0 hr 25 minutes
Reflected power = 17.47W
Transmitted power = 108.7W
PowerSum = 126.2W
FSS:
It has been locked for 0 days 0 hr and 25 min
TPD[V] = 0.8925V
ISS:
The diffracted power is around 3.6%
Last saturation event was 0 days 0 hours and 33 minutes ago
Possible Issues:
ISS diffracted power is high