oli.patane@LIGO.ORG - posted 11:54, Thursday 07 November 2024 (81128)
Lockloss

Lockloss @ 11/07 19:31UTC, possibly due to commissioning activities. Doesn't look like anything in PSL/IMC land(ndscope1). We had some glitches in DARM the second before the lockloss, and actuation to the top mass of PRM had just been started a few seconds before, so it's possible that that was the reason(ndscope2).

Images attached to this report