david.barker@LIGO.ORG - posted 19:48, Tuesday 12 November 2024 - last comment - 11:22, Wednesday 13 November 2024(81234)
VACSTAT BSC3 single sensor glitch, no alarm issued to VAC team

At 19:27:26 PST a PT132_MOD2 glitch caused a single-sensor VACSTAT glitch. The new code requires two chambers to glitch before alarming the VAC team, so in this case no cell phone alarms were sent.

The glitch lasted 4 seconds, and as before it was DeltaP which tripped, exceeded its trip level of 1.0e-08Torr  (see attached).

I restarted vacstat_ioc.service on cdsioc0 to reset the system.

Images attached to this report
Comments related to this report
david.barker@LIGO.ORG - 11:22, Wednesday 13 November 2024 (81252)

Attached trend shows the increase in sensitity for BSC2's slope and deltap following the BSC3 glitch detection. All trip levels were later restored following a restart of vacstat_ioc.

Images attached to this comment