Displaying report 1-1 of 1.
Reports until 11:42, Tuesday 19 November 2024
H1 PSL
victoriaa.xu@LIGO.ORG - posted 11:42, Tuesday 19 November 2024 - last comment - 17:53, Tuesday 19 November 2024(81354)
PSL PMC + FSS was continuously locked over emergency vent

It looks like usually the PSL PMC + FSS can stay locked for long periods of time, for example >1 month over the emergency vent in August. See the attached trends of the # of locked days and hours in the last few months (ndscope saved as /ligo/home/victoriaa.xu/ndscope/PSL/pmc_fss_can_stay_locked_over_emergency_o4b_vent.yaml).

The last time the PMC was locked for almost a month ended on Sept 23, 2024. Since then the PSL PMC has not stayed locked for over 5 days, but this is most likely due to commissioning tests and debugging which started around then.

Images attached to this report
Comments related to this report
victoriaa.xu@LIGO.ORG - 13:20, Tuesday 19 November 2024 (81357)

Some trends showing several PSL signals over O4b, and the end of O4a.

Key points- PMC + FSS stayed locked continuously during both the O4a/b break (Feb 2024), and the emergency vent (Aug 2024), with minimal glitching in PMC TRANS and Ref Cav Trans PD (FSS-TPD).

Images attached to this comment
victoriaa.xu@LIGO.ORG - 17:53, Tuesday 19 November 2024 (81369)

Trying to compare PSL-related spectra between the original O4 NPRO and the current NPRO using DTT is kinda confusing.

Comparing times with before O4 NPRO from 21 Oct 2024 04:55 UTC (blue, 169 Mpc), and now with O3 NPRO at 16 Nov 2024 08:20 UTC (red, 169.9 Mpc).

For the fast mon spectra in the top left plot, FSS FAST MON OUT looks 2-5x noisier now than before. H1:PSL-FSS_FAST_MON_OUT_DQ calibrated into Hz/rtHz using zpk = ([], [10], [1.3e6])  (see 81251, this makes H1 and L1 spectra into comparable Hz/rtHz units 81210).

But for FSS-TPD (ref cav trans), it looks like there's some extra 1-10Hz noise, but otherwise the trans spectra might be quieter? Similarly confusing, the ISS AOM control signal looks quieter. Not a clear takeaway from just these spectra on how to compare O3/O4 NPROs.

Images attached to this comment