Recent TCS CO2Y lock loss and table work today

Camilla C., TJ

Recently, the CO2Y laser that we replaced on Oct 22 has been struggling to stay locked for long periods of time (alog81271 and trend from today). We've found some loose or bad cables in the past that have caused us issues, so we went out on table today to double check they are all ok.

The RF cables that lead into the side of the laser can slightly impact the output power when wiggled, in particular the ones with a BNC connector, but not to the point that we think it would be causing issues. The only cable that we found loose was for the PZT that goes to the head of the laser. The male portion of the SMA that comes out of the laser head was loose, and cannot be tightened from outside of the laser. We verified that the connection from this to the cable were good, but wiggling it did still introduce glitched in the PZT channel. I don't think that we've conviced ourselves that this is a problem though, because the PZT doesn't seem to glitch when the laser loses lock and instead it will run away.

An unfortunate consequence of the cable wiggling was that one of the Beckhoff plugs at the feedthrough must have been unseated slightly and caused our mask flipper read backs to read incorrectly. The screws for this plug were not working so we just pushed the plug back in to fully seat it and all seemed to work again.

We still are not sure why we've been having these lock losses lately, the 2nd and 3rd attachments show a few of them from the last day or so. They remind me of back in 2019 when we saw this - example1example2. The fix ultimately a chiller swap (alog54980), but the flow and water temp seem more stable this time around. Not completely ruling it out yet though.

Images attached to this report
Comments related to this report
thomas.shaffer@LIGO.ORG - 09:24, Wednesday 04 December 2024 (81617)

We've only had two relocks in the last two weeks since we readjusted cables. This is within its normal behavior. I'll close this FRS32709 unless this suddenly becomes unstable again. Though there might be a larger problem of laser stability, I think closing the this FRS makes sense since it is referencing a specific instance of instability.

Both X&Y tend to have periods of long stretches where they don't relock, and periods where they have issues staying locked (attachment 2). Unless there are obvious issues with chiller supply temperature, loose cables, wrong settings, etc. I don't think that we have a great grasp as to why it loses lock sometimes.

Images attached to this comment