Displaying report 1-1 of 1.
Reports until 11:48, Monday 07 October 2013
H1 IOO
sheila.dwyer@LIGO.ORG - posted 11:48, Monday 07 October 2013 - last comment - 12:36, Thursday 17 October 2013(8012)
IMC transmitted beam on IOT2L

I used thenano scan to measure a beam profile on IOT2L.  near the periscope, the vertical waist diameter is 4411um.  The horizontal beam isnot gaussian, the waist diameter fit is 4007um.  (screen shot is first image attached to this report) This was taken 3 and 1/2" from the bottom periscope mirror, the periscope height is 23 and 1/2 ", and the top periscope mirror is approx 2'8 and1/2" from the viewport.  This data is saved as IOT2L1 on the nanoscan computer.  

I moved the scan head back to  14"  behind the botttom periscope mirror, and measured 4405 um diameter vertical , the horizontal beam still is non gaussian, but the diameter is 4070 um.  (screen shot is second image attached)

This data is now transfered to http://www.ligo-wa.caltech.edu/~sheila.dwyer/NanoScanData/IO/

Images attached to this report
Comments related to this report
rodica.martin@LIGO.ORG - 12:36, Thursday 17 October 2013 (8143)

According to these alogs - 6403 and 6265 - the IMC Trans beam might be clipping at the HAM2 viewport. 

Displaying report 1-1 of 1.