TITLE: 11/27 Eve Shift: 0030-0600 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Observing at 158Mpc
INCOMING OPERATOR: TJ
SHIFT SUMMARY: 1 lockloss this shift (IMC tagged LL), DRMI struggled at first but after an IA relocking was smooth. There were FSS oscillations during initial alignment. The range has been hovering just under 160Mpc.
LOG:
Start Time | System | Name | Location | Lazer_Haz | Task | Time End |
---|---|---|---|---|---|---|
23:58 | CAL | Tony | PCal Lab | Local | Measurement | 00:04 |
01:18 | CAL | Tony | PCAL lab | Local | Check shutter | 02:10 |
This is not the FSS oscillating, it's the FSS having trouble acquiring lock; the FSS Trans PD signal (TPD) is zero (cavity unlocked) except for when the cavity is near resonance (these are the peaks seen in the TPD during this relocking period). The autolocker fails to lock so the TPD drops back to zero and the temperature search continues. The pattern seen in the PMC_HV signal is the PMC following the NPRO frequency change as the FSS autolocker slowly ramps the NPRO crystal temperature to try to lock the RefCav (this is why the NPRO temperature channel has the same shape as PMC_HV). Not clear why the RefCav doesn't want to lock as quickly with this NPRO as it did with the previous ones, but we did see this behavior upon FSS recovery last Friday (I had to lock the RefCav manually so we could move on with PSL recovery as the autolocker was taking too long). I recall seeing this behavior in the past (pre-COVID) and at the time we couldn't figure out the cause. Will look into this more after the Thanksgiving holiday.