WBSC9 ETMx Alignment
IAS: J. Oberling
SEI: J. Warner, G. Grabeel
 
We aligned the Y and Z (lateral and vertical, repsectively) positions and Yaw of the ETMx this morning via HEPI adjustments (position coordinates are local to End X):
Comments related to this report
greg.grabeel@LIGO.ORG - 17:05, Thursday 17 October 2013 (8155)

The final dial indicator values for those values.

  1 2 3 4
X 0.532 -- -- 0.623
Y 0.203 0.329 0.636 0.542
Z 0.419 0.430 0.406 0.474

(Corner 4 X value is so different because I was reading the dial indicator differently, it didn't actually change that much.)

jim.warner@LIGO.ORG - 08:07, Friday 18 October 2013 (8158)

Greg and I also did a quick measurement of the level after alignment. I didn't try to calculate a height, but the table is level to within .010" over the span of the optical table, with 3 of 4 corners being more like .005" difference. This agrees with the spec used for the test stands, so I assume this is good enough.