Displaying report 1-1 of 1.
Reports until 16:17, Wednesday 04 December 2024
H1 TCS
ryan.short@LIGO.ORG - posted 16:17, Wednesday 04 December 2024 - last comment - 10:42, Friday 03 January 2025(81622)
TCS Monthly Trends

FAMIS 28455, last checked in alog81106

Only things of note on these trends are that the ITMX spherical power has been dropping for the past week and a half, which Camilla agrees looks strange, and the ITMY SLED power has reached the lower limit of 1, so it will need replacing soon. Everything else looks normal compared to last check.

Images attached to this report
Comments related to this report
thomas.shaffer@LIGO.ORG - 07:50, Thursday 05 December 2024 (81625)

Two plots, one with the ITMX V, P, Y OSEMs in the same time frame as with Ryan's plots. This could make me believe that the spherical power is just from the normal ITMX movement. The second plot is a year+ trend and I'd say that it shows this is normal movement.

Images attached to this comment
thomas.shaffer@LIGO.ORG - 07:56, Thursday 05 December 2024 (81626)

What caught my eye with these plots is thst CO2Y power has increased in the last few weeks. We've that happen after it warms up and relocks, but this seems to be trending that way after a few relocks. Not sure how to explain that one.

Also, it looks like th flow for CO2X isn't as stable at Y, worth keeping an eye on.

Images attached to this comment
camilla.compton@LIGO.ORG - 10:42, Friday 03 January 2025 (82098)

Although ITMY POWERMON is below 1 (trend), the powermon seems to read lower than ITMX, see 73371 where they were last swapped in October 2023, both fibers had 2.5mW out but  SLEDPOWERMON recorded 5 vs 3. 

I checked that the data still makes sense and isn't considerably noisier: now vs after replacement. Maybe we can stretch out the life of the SLEDs for the end of O4 but should keep an eye on them.

You can see that the spherical power form ITMX is offset from zero so we should take new references soon.

Images attached to this comment