Reverted back dust monitor code
This falls under WP 4198.

For the time being I have reverted the dust monitor code back to device support met_one_227b-1.0.1 and IOC dust_met_one_227b-1.0.4.

I stopped the test code running for the optics labs (which I had restarted on opsws8 this morning) at 14:04.

I copied the dust target directories from /ligo/lho/h0/target to /ligo/lho/h0/target_archive. I appended _october_21_2013 to them.
I copied the dust target directories with _october_8_2013 appended to them from /ligo/lho/h0/target_archive to /ligo/lho/h0/target.

I stopped the dust IOCs running on h0epics.
I started the reverted dust IOCs on h0epics2.

Dave reverted back the INI file and restarted the DAQ.

I burtrestored the IOCs with the files in /ligo/cds/lho/h0/burt/2013/10/07/00:00.
Comments related to this report
patrick.thomas@LIGO.ORG - 15:45, Monday 21 October 2013 (8203)
Dust monitor 3 kept stopping after this change. I went out and checked the front panel. The mode was manual and the hold time was 1 second. I changed the mode to automatic and the hold time to 1 minute. While I was out there I checked dust monitor 15. It turns out that it has no fuse in the breakout box. When I came back in the IOC for the dust monitors in the LVEA was giving errors. I restarted it and burtrestored it using the file in /ligo/cds/lho/h0/burt/2013/10/07/00:00.

To do: Put fuse in breakout box for dust monitor 15.