Some days, the lock losses were due to earthquakes, wind and ground motion, some due to HAM issues and some have unknown causes.
Average range was between 155-165 Mpc for the days that were majorily in locked mode.
The week had pretty high wind and ground motion for the majority part of multiple days. Some days seismic activity was also high contributing to mutiple locklosses.
The Corner BSC 10 has constant high amplitude low frequency noise multiple days.
H1 EX and H1 EY Acoustic PEM somedays has long periods of low frequency noise in the spectrograms.
Some HVETO round winner channels were new while some repeated for multiple days. See the shift summary for more details.
FScan plots were not available for 3 days of the week. While, the lines remained under 600 for the rest of the days.