IM4 trans calibration check

Sheila, Mayank

We wanted to double check the calibration of IM4 trans into power on PRM, similar to 63812 and 62213

Comments related to this report
elenna.capote@LIGO.ORG - 17:20, Friday 31 January 2025 (82571)

I have updated the IM4 trans calibration with this additional factor of 0.9566. I engaged FM8 in the IM4 trans nsum filter bank, which is now labeled "alog82260".

This indicates that at 60.1 W input power from the PSL, we have 56.6 W of power on the PRM.

Screenshots show the new filter in IM4 trans NSUM, SDF in safe, and SDF in observe.

Sheila will add her code shortly.

Images attached to this comment
sheila.dwyer@LIGO.ORG - 16:30, Monday 03 February 2025 (82613)

code used to check this calibration is attached