PSL Status Report (FAMIS #26349)

Laser Status:
    NPRO output power is 1.842W
    AMP1 output power is 70.11W
    AMP2 output power is 137.6W
    NPRO watchdog is GREEN
    AMP1 watchdog is GREEN
    AMP2 watchdog is GREEN
    PDWD watchdog is GREEN

PMC:
    It has been locked 30 days, 21 hr 54 minutes
    Reflected power = 25.75W
    Transmitted power = 102.5W
    PowerSum = 128.2W

FSS:
    It has been locked for 0 days 2 hr and 54 min
    TPD[V] = 0.773V

ISS:
    The diffracted power is around 4.0%
    Last saturation event was 0 days 2 hours and 54 minutes ago


Possible Issues:
    PMC reflected power is high