J. Kissel, L. Dartez, E. Goetz In process of updating the calibration after installing the extra AA 65k to 16k digital AA filter we turned on this morning (see 82404, 82412 and 82413), we've updated the "template" pyDARM DARM model parameter set that is the basis for all copies for every report against which the model is compared against measurement, and from which the calibration pipeline's model is derived. The changes are relatively simple, -omc_filter_noncompensating_modules = 9,10 : 9,10 +omc_filter_noncompensating_modules = 8,9,10 : 8,9,10 -omc_filter_file = Common/H1CalFilterArchive/h1iopomc0/H1IOPOMC0_1364929770.txt +omc_filter_file = Common/H1CalFilterArchive/h1iopomc0/H1IOPOMC0_1421610658.txt where the "-" lines are the "before" and the "+" lines are the "after." Here's the location of the file, and the corresponding "before" vs. "after" git commit hash. /ligo/groups/cal/H1/ifo/pydarm_H1.ini Previous version f480b0a1 Now new version 17649002