PR2 moves today out of lock during Commissioning

Sheila, Matt T, Mayank, Jennie W

 

tl;dr: After previous PR2 moves last week, we measured the spot clipping on scraper baffle (and exiting HAM3 viewport) to be 8mW in lock. After measuring this we unlocked for unknown reason and so team PR2 moved 6mm towards the centre of PR2 in yaw while unlocked. We estimate we are roughly centred on PR2 now.

Today Sheila and I measured the spot coming out of the HAM3 viewport as was done by Robert in this entry. The process is replace the black glass guillotine with lexan guillotine, take off the illuminator, clip a lens on a stalk to the top of the guillotine and align this with the beam coming out of the viewport. Not all the light is now captured by the lens as the spot is now more a semi-circle of a clipped beam. We moved the power metre behind the lens till we found the max power we could get on the power metre. This was 8mW and the PR3 yaw slider was at around -74 microradians.


After we came back to control room the IFO lost lock so it was decided to move towards the centre of PR2 which had been measured by Sheila using an A2L measurement at the start of the commissioning period.

Matt tuned down on PR3 yaw in 3 microradian steps to -232.4 microradians. Each step he tuned the 8X picomotor to bring back the DIFF and COMM ALS beatnotes. Every so often it is also neccessary to step up the pitch on PR3 to compensate for the cross-coupling of yaw to pitch. The ndscope image for the moves is here. Matt left the picomotor like this and the sliders like this.

Images attached to this report
Comments related to this report
sheila.dwyer@LIGO.ORG - 13:06, Monday 10 February 2025 (82725)

Before today's move PR3 yaw slider was at -74, where we measured the Y2L coefficent to be -3, which corresponds to 6 mm. 

We moved using the procedure in 82670, and we decided to move to -230urad on the PR3 yaw slider based on the table in 82688

We did go to ISCT1 ad fix up the beatnotes, and TJ stopped at PREP_DC_READOUT where I pico'd to center on the POP QPDs.  The centering actually could be better, with POP A well centered POP B is at -0.5 in pitch and -0.1 in yaw at 2W with all ASC on.  We can fix this another day, but didn't want to take the time today.

I've set the POP A offsets to 0, accepted this in safe.snap, and added PRC1 and PRC2 in the guardian.