Displaying report 1-1 of 1.
Reports until 07:41, Wednesday 12 February 2025
LHO General
thomas.shaffer@LIGO.ORG - posted 07:41, Wednesday 12 February 2025 - last comment - 10:22, Wednesday 12 February 2025(82762)
Ops Day Shift Start

TITLE: 02/12 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Lock Acquisition
OUTGOING OPERATOR: Oli
CURRENT ENVIRONMENT:
   SEI_ENV state: CALM
   Wind: 3mph Gusts, 1mph 3min avg
   Primary useism: 0.02 μm/s
   Secondary useism: 0.34 μm/s
QUICK SUMMARY: Lost lock just before I got in. Looks like ALS has been been unable to stay locked even with good buildup, something we've seen over the last week or so. Running an initial alignment will generally fix this, and so far it has. VEA temperatures are not very stable, but Eric is looking into it.
 

Comments related to this report
ryan.crouch@LIGO.ORG - 10:22, Wednesday 12 February 2025 (82767)SUS

ITMY mode8's damping wasn't going well this morning, it was slowly ringing up, I'm guessing its due to the slight temperature excursions in the LVEA. I turned off the -30 phase filter and used a positive gain and it started to damp, I'll put the gain to zero in lscparams just in case. I reloaded VIOLIN_DAMPING after we lost lock so the changes should be live.

So the new settings I've found are FM1 + FM10, G = = +0.3

Displaying report 1-1 of 1.