TITLE: 02/22 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 153Mpc
OUTGOING OPERATOR: Oli
CURRENT ENVIRONMENT:
SEI_ENV state: USEISM
Wind: 3mph Gusts, 0mph 3min avg
Primary useism: 0.03 μm/s
Secondary useism: 0.45 μm/s
QUICK SUMMARY:
Nice shift for H1 with a lock over 12.5hrs. Also ran the Saturday calibration. Microseism is still high.
LOG:
- PSL Dust Alarms have been alarming throughout morning.
- 1847 Changed ITMy MODE-5 to Rahul's newest settings after chatting with him: FM6 + FM8 + FM10 Gain +0.01 (+30deg phase)
- BUT: This still is not set in Guardian! Guardian still uses the older settings (60degphase + a gain of 0.0). If Rahul is happy with the new settings above, we will need to:
- update lsc_params &
- Load guardian(s): would this be VIOLIN_DAMPING only or also ISC_LOOK as well??
- 1930 Calibration Time (but first---wanted to see if taking VIOLIN DAMPING to SIMPLE bumps out of OBSERVING)
- Taking VIOLIN DAMPING from DAMP_VIOLINS_FULL_POWER to DAMPING_ON_SIMPLE does in fact take H1 OUT of OBSERVING!
- Restored nominal gain for IY mode1
- 1947 SEI_ENV transitioned to CALM (had been at USEISM since 447utc)
- 2029 PI31 alert, but only for a few seconds.
- 2130-2151 PCal Lab work (tony)
- 2317-2327 Grabbing an optic & checking terminte status in the optics lab (robert)