Displaying report 1-1 of 1.
Reports until 08:53, Monday 24 February 2025
H1 ISC (ISC)
corey.gray@LIGO.ORG - posted 08:53, Monday 24 February 2025 - last comment - 09:25, Monday 24 February 2025(82997)
Odd Noisy Behavior for OFFLOASD DRMI ASC State

After a 2nd alignment and finally getting DRMI to lock (post-wind storm), OFFLOAD_DRMI_ASC had a very odd behavior with POP90 & PR_GAIN getting very LOUD (see attachement#1) and for ASC, the SRC2 pitch and yaw also had very odd and noisy behavior (see attachement #2).  There were also SRM & SR2 verbal alarms as well.

Stopped at OFFLOAD_DRMI_ASC as soon as we noticed this, and then it cleared up after about 2-minutes.

Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 09:11, Monday 24 February 2025 (82998)

Sheila, Jennie, Camilla.  We checked SRM and SR2 and it doesn't seem like it's a single coil driver that's causing an issue, plot attached.

Images attached to this comment
jennifer.wright@LIGO.ORG - 09:25, Monday 24 February 2025 (82999)

Shown below are plots of an oscillation starting in SRC2 that seems to begin one of thse noisy states in OFFLOAD_DRMI_ASC (111) which seems to be about 0.97 Hz. This also happened about 19:57 UTC on Sunday, as well as this morning as discussed above.

Is SRC2 going unstable at this frequency?

 

Images attached to this comment
Displaying report 1-1 of 1.