Tue Feb 25 10:17:54 2025 INFO: Fill completed in 17min 49secs
Today was the first run of the new code, which terminates the fill on rate of thermocouple temperature drop rather than dropping below a set limit. This will free us from having to manage the limit during seasonal changes.
TCmins[-144C, -140C] OAT (+11C, 51F).
For now the plot still shows the old -90C trip level. I'm working on a new plot which shows the rate-of-change channels.