SQZ NLG Data Set

Sheila, Camilla

NLG Measurements (follow 76542):

opo_grTrans_ setpoint_uW Amplified Max Amplified Min UnAmp Dark NLG (usual) NLG (maxmin) OPO Gain
120 0.05615 0.000255 0.000946 2.5e-5 61.0 62.7 -12
110 0.03241 0.000273     35.2 35.4 -11
90 0.01494 0.000288     16.2 16.8 -9
70 0.008177 0.000319     8.9 9.2 -7
40 0.003717 0.000392     4.0 4.2 -4
25 0.002531 0.000464     2.7 2.8 -2
80 0.010866 0.000302     11.8 12.2 -8

For 25uW we needed to reduce H1:SQZ-OPO_PZT_1_SCAN_TRIGGER_CHANNEL_1_LEVEL from 0.5 to 0.25

Data saved to camilla.compton/Documents/sqz/templates/dtt/20250314_NLG.xml
Example of data attached. Interestingly when we took the intal NO SQZ data, DARM looks different than in 2024, attached, using H1:CAL-DELTAL_EXTERNAL_DQ

FC2 is misaligned during this dataset. Aimed for 3m30 of data for each point.

In the Mean SQZ data, there was a strange peak around 300Hz, we checked if tis was seeding by moving H1:SQZ-LO_SERVO_SLOWOUTOFS, this made no difference

Type Time (UTC) NLG SQZ dB  @ 1kHz Angle DTT Ref
No SQZ 16:18:30 - 16:22:30 N/A   N/A ref 0
ASQZ 16:34:30 - 16:38:00 61 21.2 (-)86 ref1
ASQZ -10deg 16:38:30 - 16:42:00 61 20.1 (-)75 ref2
SQZ 16:45:00 - 16:48:30 61 -4.5 166 ref3
MSQZ 16:49:00 - 16:52:30 61 18.6 N/A ref4
ASQZ 17:04:30 - 17:08:00 35 19.2 (-)88 ref5
SQZ 17:11:30 - 17:15:00 35 -4.6 170 ref6
MSQZ 17:15:30 - 17:19:00 35 16.3 N/A ref7
ASQZ 17:31:00 - 17:34:30 16 15.4 (-)93 ref8
SQZ 17:38:00 - 17:41:30 16 -4.6 157 ref9
MSQZ 17:42:00 - 17:45:30 16 12.4 N/A ref10
ASQZ 17:55:30 - 17:59:00 9 12.5 (-)96 ref11
SQZ 18:03:00 - 18:06:30 9 -4.5 161 ref12
MSQZ 18:07:00 - 18:10:30 9 9.5 N/A ref13
ASQZ 18:18:30 - 18:22:00 4 8.2 (-)114 ref14
SQZ 18:25:00 - 18:28:30 4 -3.9 155 ref15
MSQZ 18:29:00 - 18:32:30 4 5.5 N/A ref16
ASQZ 18:48:00 - 18:51:30 2.7 5.9 (-)115 ref17
SQZ 18:52:30 - 18:56:00 2.7 -3.4 155 ref18
MSQZ 18:57:30 - 19:01:00 2.7 3.6 N/A ref19