Reduced HAM7 rejected pump power and increased SHG launch, turned OPO trans setpoint up to 120uW and measured NLG with 76542 to be 58 (this was a little lower than with 120uW in 83370).
Data attached with filename shown on screenshot.
| Type | Time (UTC) | SRCL | Angle | DTT Ref |
| No SQZ | 16:01:00 - 16:15:00 | N/A | N/A | ref 0 |
| FIS | 16:20:30 - 16:23:30 | -191 | (CLF-) 174 | ref1 |
| FIS | 16:27:30 - 16:30:30 | -90 | (CLF-) 218 | ref2 |
| FIS | 16:35:00 - 16:38:00 | 0 | (CLF+) 107 | ref3 |
| FIS | 16:42:00 - 16:45:00 | -290 | (CLF-) 146 | ref4 |
| FIS | 16:48:00 - 16:51:00 | -390 | (CLF-) 129 | ref5 |
Note that I left the OPO servo gain at -8, but we have previously used -12dB for 120uW OPO trans (83370)
| opo_grTrans_ setpoint_uW | Amplified Max | Amplified Min | UnAmp | Dark | NLG (usual) | NLG (maxmin) | OPO Gain |
| 120 | 0.0540944 | 0.00026378 | 0.000913452 | -0.0000233 | 57.75 | 58.68 | -8 |