While the SQZ team was commissioning in the control room, the HAM7 ISI tripped 4 times on the CPSs. Looking at the trip plots this looks like the CPS glitches that we have seen in the past (attachment 1).
HAM7 definitely trips more forequently than other ham chambers (attachment 2), and we know that we have seen these when there are people working in the racks nearby, but today the sqz team was adjusting TEC temperatures, closing beam diverters, or just doing nothing when these happened.
FRS33714 filed since I couldn't find one specifically for this, though we do have tickets open for other HAM6 CPS gltiching: FRS: 11640 FRS: 11642 Aggregate IIET: 11643.