Reports until 09:02, Wednesday 02 April 2025
H1 PEM
ryan.crouch@LIGO.ORG - posted 09:02, Wednesday 02 April 2025 (83694)
HAM1 dust monitor

Yesterday I moved a pumped dust monitor into the HAM1 clean room (dust monitor LVEA5). I set the alarm levels to clean-100 following E1600132 and took a look at the trend overnight. The 0.5u particle count was steady around zero and the 0.3u was around ~10 counts overnight.

Images attached to this report