Ibrahim, Ryan S, Erik, Dave:
Around the same time as there was work being done at the corner station mechanical room mezzanine, h1sush7 had a timing glitch requiring the restart of the h1iopsush7 model (and therefore all user models).
Ibrahim and Ryan S consolidated the SDFs for h1susfc1, h1sussqzin and verified the mezanine work was completed.
We restarted all the models, reset the SWWD, cleared IPC and CRC errors to return the system to nominal.
I have added this event to an already opened ticket for h1sush7 FRS20317
Wed02Apr2025
LOC TIME HOSTNAME MODEL/REBOOT
09:55:21 h1sush7 h1iopsush7
09:55:35 h1sush7 h1susfc1
09:55:49 h1sush7 h1sussqzin
09:56:03 h1sush7 h1susauxh7