Closes FAMIS 26397, last checked in alog 84018
Laser Status:
NPRO output power is 1.841W
AMP1 output power is 70.43W
AMP2 output power is 140.4W
NPRO watchdog is GREEN
AMP1 watchdog is GREEN
AMP2 watchdog is GREEN
PDWD watchdog is GREEN
PMC:
It has been locked 2 days, 23 hr 11 minutes
Reflected power = 23.28W
Transmitted power = 105.5W
PowerSum = 128.8W
FSS:
It has been locked for 1 days 20 hr and 35 min
TPD[V] = 0.8003V
ISS:
The diffracted power is around 3.7%
Last saturation event was 1 days 20 hours and 35 minutes ago
Possible Issues:
PMC reflected power is high