FAMIS 31081 (this also serves as a checkup on the PSL following the large power outage early last week)
Generally things are behaving as expected with just a few exceptions. Most notably, both amplifier output powers are ~1W lower than before the outage, likely due to the current being lower on the pump diodes in AMP1 by ~0.1A each. We had set the currents for AMP1 (power supplies 1 and 2) back to pre-outage levels of 8.7A, but perhaps due to rounding issues coming out of Beckhoff (?) the actual current is apparently lower. We may bump these up by another tenth of an amp to see if this brings the output power back up.
PMC transmitted power is lower by ~1W (and noisier), either due to the lower power out of the amplifiers or the misalignment I can clearly see on the PMC REFL camera. I'll plan to try and touch that up, along with RefCav alignment, before the work in the enclosure for the installation of the SPI pickoff starts tomorrow. Edit: I forgot that the picomotor driver for the PSL is shared with IOT2L and lives on that table which is fully disconnected for vent work.