Reports until 13:47, Monday 14 April 2025
H1 ISC (TCS)
camilla.compton@LIGO.ORG - posted 13:47, Monday 14 April 2025 (83901)
Checking Return EY ALS beam with QPD offsets on and off.

TJ, Sheila, Matt, Andrew (job shadow today), Camilla. WP12433. Table layout D1400241.

Following on from other beam profile work at the end stations 81358, Madi Simmonds models G2500646 and the ideas we had for understanding ETM mode matching better (googledoc), today we took some beam profiles of the EY ALS beam reflected of ETM HR in the HWS path.

We added a 532nm quarter waveplate downstream of the beamspiltter (ALS-M11) and tuned it to maximize the light reflected into the HWS path. This was removed when we were finished and left with Nanoscan beam profile.

We first took some beam profile measurements (photos 1-4) first with the QPD offsets on nominally, and then with the offsets zero'ed (phots 5-9). Unless the beam has been pico-ed (which it may have) the zero offsets should have the beam more centered on the Transmon parabolic mirrors. The beam changed when the offsets were turned off but did not look better/worse, just different. Compare photo 4 (QPD offsets on) and photo 5 (offsets off).

Photo
#
 
A1 horizontal
13.5%
A2 vertical
13.5%
A1 horizontal
50%
A2 vertical
50%
Started with QPDs on at thier usual values:
1
50.5cm downstream of HWS-M3 (between M3 and M4)
1485
1685
723
871
2
42.5cm downstream of HWS-M3 (between M3 and M4)
1783
1985
877
970
3
36.5cm downstream of HWS-M3 (between M3 and M4)
1872
2146
924
1135
4
13.0cm downstream of HWS-M3 (between M3 and M4)
2540
2991
1247
1516
Then we turned off QPD offsets and retook data:
9
51cm downstream of HWS-M3 (between M3 and M4)
1370
1570
676
853
8
33.5cm downstream of HWS-M3 (between M3 and M4)
1939
2184
959
1086
7
48.5cm downstream of HWS-M3 (between M3 and M4)
1440
1655
706
876
6
36cm downstream of HWS-M3 (between M3 and M4)
1830
2130
913
1098
5
13.0cm downstream of HWS-M3 (between M3 and M4)  
2186
2861
1045
1077
Images attached to this report